Frequently asked questions (FAQ) about the EUCAST reference method for MIC testing of the Mycobacterium tuberculosis complex and the SOP for calibrating surrogate methods.
The media for the reference method should be prepared from the powder base which is supplied by the manufacturer.
As stated in the reference method, 7H9 needs to be prepared according to the manufacturers’ instructions and should thus also be stored according to the instructions of the manufacturer. It is not an additional requirement to use fresh 7H9 as long as the media is handled according to the instructions of the manufacturer.
The OADC should be added according to the instructions from respective manufacturer.
OADC can be purchased pre-made. 7H9 or OADC does not need to be produced from each component but may be purchased ready-made and then prepared, used and stored according to the instructions of the manufacturer.
As long as the reference method description is adhered to, it is possible to use several inoculum preparations from the same isolate on the same plate.
As long as the plate outline is adhered to (Appendix 1), several drugs can be tested but as there has to be an inoculum control for each isolate (GC1% and GC100%) it may not be suitable to use more than one isolate per plate.
As outlined in the SOP, there has to be two separate laboratories.
As outlined in the SOP, there is a need to check for media variability for each drug as the potential media-drug variability is specific to each drug.
It should be 3 separate McF preparations (individual inocula).
It is correct according to the calibration SOP. Stage 1 is to ensure that there is no technical issue in the drug-media combination which would make the following testing problematic and needs to be solved first. In stage 3, two different media are needed to include any media specific variability when the ECOFF is determined.
The rule about testing 8 drug concentrations applies to the reference method. As long as there is no truncation, it is possible to use 7 tubes for the comparator method. If needed, two ranges of MIC values for the same drug may be used (low and high).
The same drug as used in the comparator method in the calibration SOP should not be used (ie lyophilised drugs for MGIT tubes produced by BD).
It is important that the inoculum for the reference method is prepared according to the protocol. The preparation of the inoculum in the comparator method is up to the manufacturer to decide. However, it must be prepared in a reproducible way.
Download pdf version
Looking for more information? Get in touch with an EUCAST representative today!
We’ve prepared a short video to help you get familiar with the new layout and features.